IM DA100-11E
7-10
Data status
Outputs the status of the report computation result. The sum of the top numbers of the phenomena
from the following items is output.
1H: Detected over-limit data during measurement
2H: Detected special data during sampling
4H: Power failure occurred during computation
80H: Numerical data valid
Output data
Depending on the type of report computation, the results are output in the following order.
Computation type: INST (instantaneous value)
G1 to Gn, data H1 to data Hn: Instantaneous value during report creation
l1 to ln, data J1 to data Jn: Meaningless data
K1 to Kn, data L1 to Ln: Meaningless data
Computation type: AVE (average)
G1 to Gn, data H1 to data Hn:
Hourly report: Average value of the measured data over 1 hour
Dayly report: Average value of the measured data over 1 day
Monthly report: Average value of the measured data over 1 month
l1 to ln, data J1 to data Jn:
Hourly report: Minimum value of the measured data over 1 hour
Dayly report: Minimum value of the measured data over 1 day
Monthly report: Minimum value of the measured data over 1 month
K1 to Kn, data L1 to Ln:
Hourly report: Maximum value of the measured data over 1 hour
Dayly report: Maximum value of the measured data over 1 day
Monthly report: Maximum value of the measured data over 1 month
Computation type: SUM (sum)
G1 to Gn, data H1 to data Hn:
Hourly report: Summed value of the measured data over 1 hour
Dayly report: Summed value of the measured data over 1 day
Monthly report: Summed value of the measured data over 1 month
l1 to ln, data J1 to data Jn: Total of the summed values since the start of the report
K1 to Kn, data L1 to Ln: Meaningless data
If the data is abnormal, the following value is output at the mantissa of the computed data.
7FFF7FFFH: Positive over-limit data
80018001H: Negative over-limit data
80028002H: Measurement range setting skips
80038003H: The specified channel is not connected
80048004H: Data error
80058005H: Data output not possible
Note
The mantissa of the data length, extension number, power failure information and data can be output from either the upper
or the lower byte with the BO command.
The report channels set to OFF are output in the following way.
Type of report computation result: 00H (invalid)
Data status: Meaningless data
The exponent and mantissa of the data: Meaningless data
7.9 Report Output Format