IM DA100-11E
5-4
5.3 Setting a Channel No., and Alarm Output Relay
No.
Channel and relay numbers are expressed as shown below in three characters.
Unit No. + Slot No. + Number in slot
1
2 3
Channel/relay No.
Slot No. (In case of channel/relay No. 10, Slot No. + 1)
Unit No. ("A" for optional computation channel)
Channel Number
Stand-alone model(DA100-1) Expandable model(DA100-2)
Unit No. 0 0 to 5 (Subunit: DS400/DS600)
Slot No. 0 to 5 0 to 5
Channel No. 1 to 10 (CH10: 0)* 1 to 10 (CH10: 0)*
*: For CH10, the channel number digit is expressed by 0 and the slot number digit, by slot number + 1.
Note
If successive channel numbers are to be specified, enter as ABC-DE, where
ABC: the above 3-digit channel number (unit no., slot no., and channel no.)
DE: the channel number except the unit number (slot no., and channel no.)
Successive channels can only be specified for a unit.
The input module cannot be mounted to the DA100 main unit of an extended model.
Relay Number
Stand-alone model(DA100-1) Expandable model(DA100-2)
Unit No. 0 I (DA100 main unit)
0 to 5 (Subunit: DS400/DS600)
Slot No. 0 to 5 0 to 5
Relay No. 1 to 10 (Relay No. 10: 0)* 1 to 10 (Relay No. 10: 0)*
*: For relay No. 10, the channel number digit is expressed by 0 and the slot number digit, by slot number + 1.
Example
Channel 9 of the module mounted in slot 2 of the unit 3 of the extended model: 329
Relay of No. 10 of the module mounted in slot 2 of the unit I of the extended model: I30
POWER
100-240V 50/60Hz 55VA MAX
DATA ACQUITION UNIT
STATUS
Slot 0
Slot 1
Slot 2
Slot 3
Unit I
(DA100)
POWER
100-240V 50/60Hz 70VA MAX
STATUS
Slot 0
Slot 1
Slot 2
Slot 3
Slot 4
Slot 5
Unit 0
0 to 5
(DS400/DS600)
SUB UNIT
POWER
100-240V 50/60Hz 70VA MAX
STATUS
Slot 0
Slot 1
Slot 2
Slot 3
Slot 4
Slot 5
SUB UNIT
POWER
100-240V 50/60Hz 70VA MAX
DATA ACQUITION UNIT
STATUS
Slot 0
Slot 1
Slot 2
Slot 3
Slot 4
Slot 5
Unit 0
Stand-alone model
Expandable model
Module
1
3
5
7
9
2
4
6
8
10
2
4
6
8
10
1
3
5
7
9
*1
*1 : The unit number of the sub-unit is the number
set by the switch located on each unit below
slot 2.
Maku sure the settings conform the switch
setting.
Unit No.
setting switch
Vue de la page 45