IM DA100-11E
6-12
If p2 is set to OFF, the relays with the setting
parameters of 1 are turned OFF, and those of 0
maintain the current condition.
Always set p3 with four characters.
The settings of bits 11 to 16 of p3 are ignored.
For the four-point alarm output relay module,
bits 5 to 16 are ignored.
6.3 Data Output Request Command
6.3 Data Output
Request Command
TS Selects the output data.
Mode All modes (A/D calibration data output is limited
to the A/D calibration mode only.)
Setting TSp1<terminator>
p1 0 Measured data output
1 Setting data output
2 Unit data output
4 Report data output
5 System configuration data output
8 A/D calibration data output
9 Setting data output in setup mode
Comments Setting for p1=0, 1 or 2 is effective only in the
operation mode.
Setting for p1=8 is effective only in the A/D
calibration mode.
The setting, p1=4 is effective when the
instrument has the report function and one of
hourly, daily and monthly reports is ON.
FM Selects the output format for measured
data.
Mode Operation mode
Setting FMp1, p2, p3<terminator>
p1 0 Outputs measured data in ASCII format.
1 Outputs measured data in binary format.
2 Outputs computed data in ASCII format.
3Outputs computed data in binary format.
p2 First output channel (001 to 560)
First output computation channel (A01 to
A60)
p3 Last output channel (001 to 560)
Last output computation channel (A01 to
A60)
Comments Be sure to specify data to be output with the TS
command and execute “GET” or “ESC T”
before transmitting the FM command.
The command selects the output format of
measured data, either ASCII or binary, and the
output channel.
If there is no input channel recognized by the
DA100 among specified channels, a syntax
error occurs.
For stand-alone models computaion channel
number is A01 to A30.
RF Selects the output format of the report.
Mode Operation mode
Setting RFp1, p2, p3<terminator>
p1 0 Output hourly report data
1 Output daily report data
2 Output monthly report data
3 Output the status of the hourly/daily/
monthly data
p2 First channel for output (R01 to R60)
p3 Last channel for output (R01 to R60)
Comments This setting is effective when the instrument
has the report function.
Before transmitting the RF command, be sure
to specify the data to output using the TS
command and execute “GET” or “ESC T.”