IM DA100-11E
6-7
Commands
6
Integration time: Equivalent to 50/60 Hz
Input module Filter off Filter on
10CH 0.5* 3
20CH 2 4
30CH 2 4
*: 2s if the power monitor module is installed
Integration time: 100 ms
Input module Filter off Filter on
10CH 4 12
20CH 5 15
30CH 6 20
XI Sets the A/D integration time.
Mode Setup mode
Setting XIp1, p2<terminator>
p1 Subunit (0 to 5)
p2 Integration time (AUTO, 50 Hz, 60 Hz,
100 ms)
Example Set the A/D integration time in subunit 0 to 50
Hz.
XI0, 50Hz
XQ Sets the filter.
Mode Setup mode
Setting XQp1<terminator>
p1 ON Filter on
OFF Filter off
Example Insert a filter in the input.
XQON
XA Performs the setting related to alarm.
Mode Setup mode
Setting XAp1, p2, p3<terminator>
p1 Interval for limit of increasing rate-of-
change (1 to 15)
p2 Interval for limit of decreasing rate-of-
change (1 to 15)
p3 Alarm hysteresis (0.0 to 1.0)
Example Set the interval for limit of increasing rate-of-
change to 10, the interval for the limit of the
decreasing rate-of-change to 10 and alarm
hysteresis to 0.5%.
Comments This command sets an interval to detect the rate-
of-change alarm and hysteresis.
Set an interval using the measurement period
as the unit.
Set the hysteresis with the percentage of the
measurement range or scaling range.
The computation channel hysteresis is fixed to
0.
XY Sets the re-alarm for a re-failure.
Mode Setup mode
Setting XYp1, p2<terminator>
p1 Reflash number (1 to 6)
p2 Off, subunit (I, 0 to 5) + relay number
Example Set reflash relay 1 to relay module 1 connected to
slot 0 of the main unit.
XY1, I01
XN Sets AND/OR of the alarm output relay.
Mode Setup mode
Setting XNp1, p2<terminator>
p1 Unit
I: Main unit (DA100)
0 to 5: Subunit
p2 Relay number to be set to AND (01 to 60)
NONE: All are OR.
01-XX (XX=01 to 60)
Relays whose numbers are 1 to XX are
set to AND.
Example Have No. 1 of slot 0 to No. 10 of slot 2 in the
alarm output relay module connected to the main
unit be AND relay outputs.
XNI, I, 01-20
Comments Setting is effective only in the same unit.
Set p2 with the first relay number (01=fixed)
to the last relay number. If all OR output is to
be set, set p2 to NONE.
An error will occur when alarm output module or
DI/DO module isn’t mounted to selected DA100
or DS400/DS600.
XD Sets energizing or deenergizing.
Mode Setup mode
Setting XDp1, p2<terminator>
p1 Unit (I, 0 to 5) + Relay number (01 to 60)
p2 Selection of energizing/deenergizing
ENERG Energizing
DE-EN Deenergizing
Example Set the relays of No. 1 in slot 0 to No. 10 in slot 5
of the relay module connected to subunit 5 to
energizing outputs.
XD501-60, ENERG
Comments If a successive number of output relays are to
be set, insert a minus sign (-) between the first
relay number and the last relay number, as
shown in the example. However, relays that
can be successively set are those in the same
subunit.
An error will occur when alarm output module or
DI/DO module isn’t mounted to selected DA100
or DS400/DS600.
XH Sets the hold/non-hold.
Mode Setup mode
Setting XHp1<terminator>
p1 Output relay hold on/off
ON Hold
OFF Non-hold
Example Hold the output state of an alarm output relay.
XHON
XB Sets the burnout.
Mode Setup mode
Setting XBp1, p2<terminator>
p1 Channel number (001 to 560)
p2 Selection of burnout
OFF
UP Upscale travelling beyond
the scale
DOWN Downscale travelling
beyond the scale
Example Set channels 01 to 10 of subunit 0 to upscale
burnout.
XB001-10, UP
Comments If channels are to be set successively, the setting
is effective only when they are in the same unit.
6.1 Setting Command