IM DA100-11E
7-6
7.6 Output Format for Unit and Decimal Point
Position
These outputs are issued in the following format by the TS2 + “Device Trigger (GET)” + LF command.
S1S2CCCUUUUUU, PCrLf
Each symbol denotes the following:
S1: Data status 1
N --------------------- Normal
D --------------------- Differential input
S ---------------------- Measurement range skips.
S2: Data status 2
Space ---------------- Interim data
E ---------------------- Final data
CCC: Channel number (3 characters)
UUUUUU: Unit (6 characters)
mV
------- mV
V
------ V
C ------ °C
P: Decimal point position (0 to 5)
0 ---------------------- 00000
1 ---------------------- 0000.0
2 ---------------------- 000.00
3 ---------------------- 00.000
4 ---------------------- 0.0000
Note
• Data in the channels not connected in the system settings, including channel numbers, are not output.