Yokogawa DA100 Manuel d'utilisateur Page 74

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 105
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 73
IM DA100-11E
7-9
Output Format
7
7.9 Report Output Format
Hourly report
The hourly report is output in the following format with the command, TS4+device
trigger(GET)+RF0.
Data
length
Year Month
Day Hour
Minute
0
A1
B1
C1 D1
An Bn Cn Dn
Number of output bytes
Time when the report is created
First channel
Last channel
A1 to An: Report channel number ( 00H to 3BH)
B1 to Bn: Unit number (Fixed to 80H during computation)
C1 to Cn: Channel number*
D1 to Dn: Report computation type
*: Computation channels are
00H to 1DH for stand-alone type
00H to 3BH for extended type
0
Extension
number
Power failure data 0: no failure 1:failure occurred
0 or 1
A1
An
A1
An
A1
An
E1 F1 G1
DATA H1
En Fn Gn
DATA Hn
E1 F1 I1
DATA J1
DATA Jn
E1 F1 K1
DATA L1
DATA Ln
En Fn In
En Fn Kn
Report computation data for each report channel
A1 ot An: Report channel number : 00H to 3BH
E1 to En:
Type of the report computation result to output
F1 to Fn: Report computation data status
G1 to Gn, l1 to ln, K1 to Kn:
Mantissa of the report computation data
Data H1 to data Hn:
Mantissa of the report computation data
Data J1 to data Jn:
Mantissa of the report computation data
Data L1 to data Ln:
Mantissa of the report computation data
Starting date and time of the sum for TOTAL
Instantaneous value, average value
or summed value
Minimum value or Total of
the summed values
Maximum value
Year Month
Day Hour
Minute
Note
RF0 outputs data on the specified report channles.
Data length
The number of output bytes can be determined from the following expression.
Number of output bytes = 16 + 4 x N + 8 x N x 3 + (8 + 8 + N) x extension number
N: Number of output channels
Report computation type
Output the type of computation set for each report channel.
00H: OFF
01H: INST (Instantaneous value)
02H: AVE (Average)
03H: SUM (Sum)
The type of report computation result to output
The report computation results vary depending on the report computation types. The following
types of report computation results are output.
Output value Type of report computation result to output Report computation type
00H Invalid data OFF
01H INST (instantaneous value) INST
02H AVE (average value) AVE
03H SUM (summed value) SUM
04H MIN (minimum value) AVE
05H MAX (maximum value) AVE
06H TOTAL (total of the summed values) SUM