5-2
5.1 Command Format
Terminator
Any one of the following forms a terminator.
CR + LF
•LF
EOI = True
(If EOI is to be used for a terminator, add EOI = True to the last parameter character.)
Sub-delimita
Several commands can be executed in a row when they are divided by a semicolon (;).
Example
XA2, 2, 0.5 ; XV4 ; XI2, AUTO CrLf
Note
The total data length from the first character to the terminator should not exceed 200 bytes.
•A sub-delimita cannot be used for the commands FM, LF, CF, RF, VF, RC, RS, BL, DS, XE, or XZ. These commands can
only be carried out one by one.
In case of RS-232-C one ACK-status (E0/E1) will be output for each command devided by ;.