Error message<Software Configuration>
No. Error/Corrective Action
E3301 Failed to scan.
Check that the DARWIN is turned ON, the communication module is being recognized by the DARWIN (DR/DC), and the
cable is connected properly.
In addition, check the following items according to the communication methods.
For GPIB
Is the address correct? Is the driver is properly installed? Is the GPIB card manufactured by NI?
For Ether
Is the address correct? Is the TCP/IP protocol installed in Windows? Is the Ether card properly installed?
For RS-232-C, RS-422-A, and RS-485
Is the baud rate correct? Is the port (COM1 to COM4) correct? Is the address correct (RS422)? Is the PC's serial port
functioning? Is the correct type of cable being used?
E3302 Failed to reconstruct
See the corrective actions for E3301.
E3303 Failed to initialize
See the corrective actions for E3301.
E3312 Select all ranges that has invalid values.
Even though there is a range with an invalid calibration value, that range is not selected. Select the range with the invalid
calibration value and calibrate it.
E3313 Failed to calibrate.
Check that the module is connected properly. Check the wiring, rated input, and etc.
E3314 For xxx is invalid.
Enter the correct value.
E3331 There is no module. Or there are many modules.
Several modules exist when setting up the network. Specify only one module.
E3337 The address is invalid.
Properly set the combination of the IP address, subnet mask, and default gateway.
E3338 Can not set the network address.
An error occurred while setting the address.
E3339 Modules not found.
Set the module's dipswitch and turn DARWIN ON.
Check that the cables are connected properly.
Error message<Historical Viewer >
No. Error/Corrective Action
E3101 Data is not exist.
Valid data do not exist.
E3102 Channel is not exist
Valid channel does not exist.