IM DA100-01E
2-16
Wiring Input Signal Lines (to Universal and DCV/TC/DI input modules)
Terminals
Screw type terminal
Clamp type terminal*
2
+
-
A
Bb
DC voltage TC
contact
RTD*
1
A
B
b
Channel1
Channel2
Channel10
Channel2
Channel4
Channel1
Channel3
Channel9Channel10
+
-
*1 There are no RTD
input terminals on
the DCV/TC/DI input
module and digital
input module.
*2 There is no clamp
type terminal for
digital input module.
Wiring Diagram
DC voltage input
Compensation lead
DC current input
Shunt resistor
Note:For 4 to 20mA input, shunt
resistance value should
be 250 ±0.1%
10* max./leadwire
Three wire resistances
should be approx. equal.
*10 max. for Pt100 and
Pt50, 1 max. for Cu10.
DC input
RTD inputTC input
DC voltage input/DI input (contact)
bAB
A
b
B
+
-
+-+-
+
-
+-
2.6 Connecting the Signal Lines