<10. Other Functions>
10-11
IM 11M12A01-02E
10.3 Parameter Initialization
Parameter settings can be initialized to the factory default settings. Initialization can be done
for all parameters or for individual parameters. The parameters that can be initialized and their
defaults are listed in Table 10.5.
<Initialization procedure>
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display.
(2) Select “Commissioning”, next “Othersthen “Defaults”. A display like Figure 10.11. appears.
(3) Select the desired item to be initialized then a display like Figure 10.12 appears.
SelectDefaults start” then initialization starts.
Display
Calibration
Blow back
mA-outputs
Alarms
Contacts
Others
All data
Enter
Defaults
F10.11E.ai
Enter
Defaults start
Cancel
Defaults Display
Figure 10.11 Defaults Display Figure 10.12 Initialization Start Display
CAUTION
Do not turn off the power during initialization. Otherwise, initialization will not be performed
properly.
8th Edition : Jan.13,2012-00