Yokogawa Integral Oxygen Analyzer ZR202 Manuel d'utilisateur Page 71

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 157
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 70
<7. Startup>
7-6
IM 11M12A01-04E
7.4.4 Selecting Parameter Codes
Switch operation Display Description
>
ENT
A01
Password has been entered and the parameter code selection display has
appeared.
Character A is ashing, indicating that character A can be changed.
>
ENT
A01
If you touch the [>] key once, the position of the digit that is ashing will move to
the right.
This allows you to change 0.
>
ENT
A01
Touch the [>] key again to move the position of the digit that is ashing to the right
one more digit. This enables you to change numeric character 1.
>
ENT
A01
Touch the [>] key again to return the position of the digit that is ashing to A.
Continuously touch the [>] key, and the position of the digit that is ashing will
move continuously to the right.
>
ENT
b01
If you touch the [] key once, character A will change to B.
>
ENT
C01
Touch the [] key once to change to C.
>
ENT
d01
Continuously touch the [] key, and the value of the digit that is ashing will
increase continuously, from D to E to F to G to A. Numeric values will change from
0 to 1 to 2 to 3 … to 8 to 9 and back to 0. However, numbers that are not present
in the parameter codes will be skipped. Each digit is changed independently. Even
though a low-order digit changes from 9 to 0, a high-order digit will not be carried.
>
ENT Set
Value
After you select the desired character, touch the [ENT] key. The set data will be
displayed.
The symbol [ ] indicates that the key is being touched. Light characters indicates that the digits are ashing.
7.4.5 Changing Set Values
(1) Selecting numeric values from among preset values
Switch operation Display Description
>
ENT
0
The set value is displayed after the parameter code selection. An example of how
to select either 0, 1, or 2 as the set value is given below.
(The currently set value is 0.)
>
ENT
1
Touch the [] key once to change the current value from 0 to 1.
>
ENT
2
Touch the [] key again to change to the numeric value 2.
>
ENT
0
If you touch the [] key again, the numeric value will return to 0.
Continuously touch the [] key, and the numeric values will change continuously.
>
ENT
C01
Display the desired numeric value and touch the [ENT] key.
The display will then return to the parameter code selection
Vue de la page 70
1 2 ... 66 67 68 69 70 71 72 73 74 75 76 ... 156 157

Commentaires sur ces manuels

Pas de commentaire