<10. Other Functions>
10-12
IM 11M12A01-04E
Calibration-related Items in Group B
Code Item Tuning Engineering unit Default setting
B01 Zero gas concentration 0.3 to 100 % O
2
1% O
2
B02 Span gas concentration 4.5 to 100 % O
2
21% O
2
B03 Calibration mode 0 Manual calibration Manual calibration
1 Semi-automatic and manual
calibration
2 Automatic, semi-automatic, and
manual calibration
B04 Hold (Output stabilization)
time
0 minutes, 0 seconds to 60 minutes,
59 seconds
MM.SS 10 minutes, 0
seconds
B05 Calibration time 0 minutes, 0 seconds to 60 minutes,
59 seconds
MM.SS 10 minutes, 0
seconds
B06 Calibration interval 0 days 0 hours to 255 days 23 hours DD.HH 30 days, 0 hours
B07 Calibration start date and
time
YY.MM.DD/HH.MM 00.01.01.00.00
B08 Calibration procedure 0 Zero and span Zero and span
1 Span only
2 Zero only
B09 Calibration concentration
measurement
Display only % O
2
B10 Manual calibration
implementation
B11 Semi-automatic calibration
implementation