Yokogawa Integral Oxygen Analyzer ZR202 Manuel d'utilisateur Page 103

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 169
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 102
IM 11M13A01-04E
8-15
8. Detailed Data Setting
8.7.2 Setting Periods over which Average Values are Calculated and Periods over which
Maximum and Minimum Values Are Monitored
The equipment enables the display of oxygen concentration average values and maxi-
mum and minimum values under measurement (see Section 10.1.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration
average values are calculated and maximum and minimum values are monitored.
8.7.2.1 Procedure
Use the parameter-code table below to set the average, maximum and minimum oxygen
concentration values. Periods over which average is calculated and periods over which
maximum and minimum values are monitored can be set, ranging from 1 to 255 hours.
If the set ranges are beyond the limits specified, an “ERR” will be displayed.
Table 8.15 Parameter Codes for Average, Maximum and Minimum Values
Set item Parameter code Set range Units
Periods over which average F11 1 to 255 Hours
values are calculated
Periods over which maximum F12 1 to 255 Hours
and minimum values are
monitored
T8.15E.EPS
8.7.2.2 Default Value
When the analyzer is delivered, or if data are initialized, periods over which average
values are calculated are set to one hour, and periods over which maximum and mini-
mum values are monitored are set to 24 hours.
Vue de la page 102
1 2 ... 98 99 100 101 102 103 104 105 106 107 108 ... 168 169

Commentaires sur ces manuels

Pas de commentaire