Yokogawa ADMAG AXR Manuel d'utilisateur Page 58

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 194
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 57
<6. PARAMETER DESCRIPTION>
6-10
IM 01E30D01-01EN
Item
Name
R
/W
Data range
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
Display unit
(BRAIN)
Display unit
/
BRAIN
G42 Alm Record Time1
(ALM TIME 1)
R 0D 00:00 to
9999D 23:59
Displays the operation time
at the occurrence of the most
recent alarm.
G43 Alm Record2
(ALM RECORD2)
R Same as G41
(Alm Record 1)
Displays the content of the
second most recent alarm.
G44 Alm Record Time2
(ALM TIME 2)
R 0D 00:00 to
9999D 23:59
Displays the operation time at
occurrence of the second most
recent alarm.
G45 Alm Record3
(ALM RECORD3)
R Same as G41
(Alm Record 1)
Displays the content of the third
most recent alarm.
G46 Alm Record Time3
(ALM TIME 3)
R 0D 00:00 to
9999D 23:59
Displays the operation time at
the occurrence of the third most
recent alarm.
G47 Alm Record4
(ALM RECORD4)
R Same as G41
(Alm Record 1)
Displays the content of the
fourth most recent alarm.
G48 Alm Record Time4
(ALM TIME 4)
R 0D 00:00 to
9999D 23:59
Displays the operation time at
the occurrence of the fourth
most recent alarm.
G60
(SELF CHECK)
R Good
Error
See6.5 Alarm Functions”.