Yokogawa PC-Based MX100 Manuel d'utilisateur Page 63

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 76
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 62
5-3
IM MW180-01E
1
2
3
4
5
App
Index
Troubleshooting
The connected input/output module is not detected.
Probable Cause Corrective Action/Description Reference Page
Module connection or module startup error.
Attached the module while the power was ON.
Turn OFF the power. Detach the input/output module
once and attach it again.
*
The PC software, main module, and input/output
module style and release number rule is not
upheld.
Check the PC software release number and main
module style number before upgrading the version.
[PC software release no.] ≥ [main module style no.] ≥
[input/output module style no.]
**
Carried out an incorrect calibration. Recalibrate. 4-2
* See the MW100 Data Acquisition Unit Users Manual (IM MW100-01E).
** See the MW100 Data Acquisition Unit Operation Guide (IM MW100-01E).
5.1 Troubleshooting
Vue de la page 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 75 76

Commentaires sur ces manuels

Pas de commentaire