Yokogawa DO402 Dissolved Oxygen Analyzer Manuel d'utilisateur Page 91

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 98
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 90
IM 12J05D02-01E
Appendix 2 11-3
3/4
QIS 12J05D02-01E
Table 3
Temperature (°C)
Resistance ()
Data Display (°C)
5 1019.5 5 ±0.3
25 1097.3 25 ±0.3
45 1174.7 45 ±0.3
3.6 Input Current Indication Check
Following Section 3.5, press the [ENT] key until the message display shows “GALVA1.”
Turn SW1 on. Change the current of the standard current generator and check the data
display. The value on the data display must be within the range shown in Table 4.
Table 4
Reference (µA) Data Display (µA)
0 0 ±0.05
8 8 ±0.05
16 16 ±0.05
Press the [ENT] key until the message display shows “GALVA2.” Change the current of
the standard current generator and check the data display. The value on the data display
must be within the range shown in Table 5.
Table 5
Reference (µA) Data Display (µA)
10 10 ±0.1
30 30 ±0.2
50 50 ±0.2
Turn SW1 off. Press the [ENT] key until the message display shows “POLAR1.” Change
the current of the standard current generator and check the data display. The value on
the data display must be within the range shown in Table 6.
Table 6
Reference (nA) Data Display (nA)
0 0 ±0.2
80 80 ±0.2
160 160 ±0.2
Press the [ENT] key until the message display shows “POLAR2.” Change the current of
the standard current generator and check the data display. The value on the data display
must be within the range shown in Table 7.
Table 7
Reference (nA) Data Display (nA)
100 100 ±0.2
300 300 ±1
500 500 ±1