Yokogawa Button Operated MV2000 Manuel d'utilisateur Page 116

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 177
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 115
4-43
Commands
M-4662
4.9 OutputCommands(RS-
422/485Commands)
ESCO Opensaninstrument
The ASCII code for ESC is 1BH. See appendix 3
for details.
Syntax
ESC
O p1<terminator>
p1
Instrument address (01 to 99)
Example Open the instrument at address 99, and enable
all commands.
ESC
O99
Description Specifies the address of the instrument to
communicate with.
You can only open one instrument.
When you open an instrument with the ESC O
command, another instrument that is currently
open is automatically closed.
When the MV receives this command
normally, the MV transmits the data “ESC O
.”
Normally, either CR+LF or LF can be used for
the terminator in communication commands.
However, the terminator for this command
must be set to CR+LF.
ESCC Closesaninstrument
The ASCII code for ESC is 1BH. See appendix 3
for details.
Syntax
ESC
C p1<terminator>
p1
Instrument address (01 to 99)
Example Close the instrument with the address 77.
ESC
C77
Description Releases the connection with the instrument.
When the MV receives this command
normally, the MV transmits the data “ESC C
.”
Normally, either CR+LF or LF can be used for
the terminator in communication commands.
However, the terminator for this command
must be set to CR+LF.
4.10 OutputCommands(Special
ResponseCommands)
*
I Transmits instrument
information
Syntax
*
I<terminator>
Description Transmits the maker, model, serial number, and
firmware version in a comma-separated ASCII
text with a terminator at the end.
Example
Omega,RD-MV1000,99AA0123,F1.01
4.9OutputCommands(RS-422/485Commands)/4.10OutputCommands(SpecialResponseCommands)