IM 12D8J2-E-E16th editionUser’s ManualModel ISC40G(S) Sensors and Fittings for Inductive Conductivity Measurement

Page 2

IM 12D8J2-E-E102.5 Operating range Conductivity : 0 – 2000 mS/cm at actual process temperature. Note: The sensor has an error (0.5 µS/cm for PEEK m

Page 3

IM 12D8J2-E-E11CSA Certificate no. : 2447837 IS, Class I Div. 1, GP A, B, C, D T4…T6 Master Contract no 182892 Electrical data : Sensor output

Page 4

IM 12D8J2-E-E12Control Drawing CSA The ISC40S sensor shall be installed withone of the Yokogawa transmitters model:• ISC202S • IC200S • FLXA21with fo

Page 5

IM 12D8J2-E-E13Control Drawing FM The ISC40S sensor shall be installed withone of the Yokogawa transmitters model:• ISC202S • IC200S • FLXA21with fol

Page 6

For optimum measurement results, the ISC40 sensor should be installed in a location that offers an acceptable representation of the process compositio

Page 7 - 1 INTRODUCTION

The sensor cable and mounting thread are pulled through the hole of the flange, and the sensor is sealed from the process by tightening the mounting n

Page 8

IM 12D8J2-E-E163.1.2 Installation of ISC40G(S)-TG with T-pieceThe ISC40G(S)-TG sensor is a PFA sensor which can be used in combination with a T-piec

Page 9 - 2 GENERAL SPECIFICATIONS

IM 12D8J2-E-E173.1.3 Installation of ISC40G(S)-GS with flange adaptersThe ISC40G(S)-GS sensor is designed for sanitary applications. For these appli

Page 10

IM 12D8J2-E-E18For mounting inISC40PR retractablefittingISC40-PRMinimum holeto fit sensorthrough: Ø48 (1.89")Ø47 (1.85")55 (2.16")ø17 (

Page 11

IM 12D8J2-E-E195 WIRING ISC40 SENSORThe ISC40 sensor is provided with a fixed cable. The connections of this dual coax cable and the Yokogawa Inductiv

Page 12 - Maximum cable

IM 12D8J2-E-E2(BG)Всички улътвания за продукти от серията ATEX Ex ce лредлагат на английски език. Ако се нуждаете oт улътвания за продукти oт серията

Page 13 - Ta 85ºC/85ºC/40ºC

IM 12D8J2-E-E206 MODELCODE ISC40 SENSORModel Code Suffix Option DescriptionISC40G General purpose inductive conductivity sensorSensor -GG

Page 14 - TORQUE MAX. 25Nm

IM 12D8J2-E-E217 SPAREPARTS ISC40 SENSORParts ISC40 sensorPart no. Description Material QuantityK1500AM Gasket Viton 5K1500AL Mounting nut AISI

Page 15 - Not recommended

IM 12D8J2-E-E228 EU DECLARATION OF CONFORMITYWe: Yokogawa Europe B.V. Euroweg 2, 3825 HD Amersfoort The Netherlandsherewith declare under our sole

Page 16 - DN80 PN10 200 160 8x ø18

IM 12D8J2-E-E239 ISC40FS – FLOW FITTING SUBASSEMBLY9.1 General Specifications9.1.1 MaterialsWetted parts Model ISC40FS-FCSA : PVDF (Kynar); Vito

Page 17

IM 12D8J2-E-E24O-ring40.64 x 5.33HolderO-ring26.57 x 3.53Ring DN50O-ring40.64 x 5.33O-ring26.57 x 3.53Ring DN50Holder88 (3.46)Ø 50Rd 78 x 1/6acc. DIN

Page 18 - 4 DIMENSIONS ISC40 SENSOR

IM 12D8J2-E-E259.4 Modelcode ISC40FSModel Suffix Option DescriptionISC40FS Flow fitting subassemblyMaterial -F PVDF -P Polypropylene -

Page 19

IM 12D8J2-E-E2610 ISC40FF – FLOW FITTING 10.1 General Specifications10.1.1 MaterialsWetted partsModel ISC40FF-S : AISI 316 SS; VitonModel ISC40FF-P

Page 20 - 6 MODELCODE ISC40 SENSOR

IM 12D8J2-E-E27Note: The ISC40 sensor type -TG does not fit in the model ISC40FF-S.10.3 Dimensions ISC40FFpipe mount (vertical)70 (2.75) pipe mount

Page 21 - 7 SPAREPARTS ISC40 SENSOR

IM 12D8J2-E-E28Fig 21: Flow fitting ISC40FF-S10.4 Modelcode ISC40FFModel Suffix Option DescriptionISC40FF flow fittingMaterial -S AISI 316 sta

Page 22

IM 12D8J2-E-E2910.5 Spareparts ISC40FFOptions ISC40FF Flow fitting, flange adaptersPart no. Description Process connection Material O-ring(s) K152

Page 23

IM 12D8J2-E-E3(LV)Visas ATEX Ex kategorijas izstrâdâjumu Lietoðanas instrukcijas tiek piegâdâtas angïu valodâs. Ja vçlaties saòemt Ex ierîèu dokumentâ

Page 24 - Ring DN50

IM 12D8J2-E-E3011 ISC40FD – IMMERSION FITTING 11.1 General Specifications11.1.1 MaterialsWetted parts without options Model ISC40FD-S : AISI 316 S

Page 25

IM 12D8J2-E-E31Fig 24: Option /FAO-Ring26.57 x 3.53O-Ring44.17 x 1.78O-Ring39.35 x 2.62ISC40FD-VISC40FD-S2-inch Stanchion (Vertical)Guide p

Page 26 - 10 ISC40FF – FLOW FITTING

IM 12D8J2-E-E32Alternative ways of mounting the immersion fitting are:• Guide pipe To facilitate this type of mounting the immersion fitting has a la

Page 27 - IC 40 FF/PVDF FIG.12

IM 12D8J2-E-E3311.4 Modelcode ISC40FDModel code Suffix code Option code DescriptionISC40FD Immersion fittingMaterial -S AISI 316 Stainless st

Page 28

IM 12D8J2-E-E3412 Chemical Compatibility Chart Material PVDF S.S. 316 VITON PEEK PP PVC PFA (Kynar)Sulfuric acid 10 OOO XXX OOO OO

Page 29

IM 12D8J2-E-E35

Page 30

IM 12D8J2-E-E Printed in The Netherlands, 16-1410 (A) ISubject to change without noticeCopyright ©IM 12X0X0-E-ESubject to change without notice Prin

Page 31

IM 12D8J2-E-E4Contents1 INTRODUCTION 71.1 General 71.2 Unpacking and Checking 71.3 Warranty and Service 71.4 Serial Number definition 82 GENERAL

Page 32 - 42 (1.65)

IM 12D8J2-E-E510 ISC40FF – FLOW FITTING 2710.1 General Specifications 27 10.1.1 Materials 27 10.1.2 Operating range 27 10.1.3 Shipping detail

Page 33

IM 12D8J2-E-E6

Page 34 - IM 12D8J2-E-E

IM 12D8J2-E-E71 INTRODUCTION1.1 GeneralThe sensor and fitting program for inductive conductivity measurement (model ISC40) is designed to meet the mo

Page 35

IM 12D8J2-E-E81.4 Serial Number definitionThe Serial Number is defined by nine (9) alphanumeric characters:X1X2 Production Location X3X4 Year/Mont

Page 36 - Copyright ©

IM 12D8J2-E-E92 GENERAL SPECIFICATIONSThe ISC40 inductive conductivity sensor is suitable for use with the Yokogawa inductive conductivity analyzers.