EJX910A and EJX930A
Multivariable Transmitter
HART Communication Type
IM 01C25R02-01E
IM 01C25R02-01E
6th Edition
Vue de la page 0
1 2 3 4 5 6 ... 102 103

Résumé du contenu

Page 1 - HART Communication Type

User’sManualEJX910A and EJX930AMultivariable TransmitterHART Communication TypeIM 01C25R02-01EIM 01C25R02-01E6th Edition

Page 2 - Contents

<2. Connection>2-3IM 01C25R02-01E(1) Conrming the device revision of the transmitter● Conrmation by using integral indicator (When the

Page 3

<Appendix2. ILBD Check List>A2-4IM 01C25R02-01EChecklist (4/5)No. Items Parameters Result Example10-a-2 L Side Blocking• Close the low-pre

Page 4

<Appendix2. ILBD Check List>A2-5IM 01C25R02-01EChecklist (5/5)No. Items Parameters Result Example10-b Simulation of Blockage detection oper

Page 5 - 1. Introduction

iIM 01C25R02-01ERevision Information Title : EJX910A and EJX930A Multivariable Transmitter HART Communication Type Manual No. : IM 01C25R02-0

Page 7 - 1.3 ATEX Documentation

<2. Connection>2-4IM 01C25R02-01E2.6 Power Supply Voltage and Load ResistanceWhen conguring the loop, make sure that the external load re

Page 8 - 2. Connection

<3. Parameter Setting>3-1IM 01C25R02-01E3.1 Menu TreeThe menu tree is different in DD(HART 5/HART 7)/DTM(HART 7) and DTM(HART 5). See the

Page 9

<3. Parameter Setting>3-2IM 01C25R02-01EAF0301-02.ai• Process variables• PV• PV % (DD) % Range (DTM)• PV AO (DD) Loop Current (DTM)• Engr

Page 10 - IM 01C25R02-01E

<3. Parameter Setting>3-3IM 01C25R02-01ESee B1(next page) BF0301-03.ai: HART 7 only• Diag/Service• Status • Test • Calibration• Diag Parame

Page 11 - Communication

<3. Parameter Setting>3-4IM 01C25R02-01EB1 • Diag ParametersF0301-04.ai• Ratio fDP Status• Ratio fDP• Ratio fSPl Status• Ratio fSPl• Rati

Page 12 - 3. Parameter Setting

<3. Parameter Setting>3-5IM 01C25R02-01E• Basic setup• Tag• Long tag• Units• Keypad input• Device infomation• OthersC• Pres Unit• SP Unit•

Page 13 - F0301-02.ai

<3. Parameter Setting>3-6IM 01C25R02-01ESee ESee D1(next page)• Detailed setup• Review• Sensors• Signal condition• Output condition• Displa

Page 14 - (next page)

<3. Parameter Setting>3-7IM 01C25R02-01ED1F0301-07.ai: HART 5 only: HART 7 onlySame as aboveSame as above• Burst Variable Code• HART output

Page 15 - F0301-04.ai

<3. Parameter Setting>3-8IM 01C25R02-01E• Detailed setup• Review• Sensors• Signal condition• Output condition• Display condition• Device in

Page 16

iIM 01C25R02-01EEJX910A and EJX930AMultivariable TransmitterHART Communication TypeIM 01C25R02-01E 6th Edition6th Edition: Mar. 2012 (YK)All Rights R

Page 17

<3. Parameter Setting>3-9IM 01C25R02-01E3.1.2 For DTM (HART 5)F0301-09.aiRoot Menu• Process Variables• Device Status• Diag and Service• Ea

Page 18 - F0301-07.ai

<3. Parameter Setting>3-10IM 01C25R02-01EF0301-10.ai• Diag and Service• Service• Advanced Diag Variables• Advanced Diag Configurations• Adv

Page 19 - F0301-08.ai

<3. Parameter Setting>3-11IM 01C25R02-01EF0301-11.ai• ConfigurationT• Pressure Sensor• Static Pressure Sensor• External Temperature Sensor•

Page 20 - 3.1.2 For DTM (HART 5)

<3. Parameter Setting>3-12IM 01C25R02-01EF0301-12.ai• ConfigurationT1• Local Display• Process Alerts• Device Information1• Device Informati

Page 21

<3. Parameter Setting>3-13IM 01C25R02-01E3.2 Basic SetupIMPORTANTAfter setting and sending data with the HART conguration tool, wait 30 s

Page 22

<3. Parameter Setting>3-14IM 01C25R02-01E• Procedure to call up the display using DD (HART 5/HART 7) and DTM (HART 7)PV related parameter[

Page 23

<3. Parameter Setting>3-15IM 01C25R02-01ENOTEThe calibration range can be set as PV LRV > PV URV under the following conditions, reversi

Page 24 - IMPORTANT

<3. Parameter Setting>3-16IM 01C25R02-01E3.2.6 Differential Pressure Signal Low Cut Mode SetupLow cut mode can be used to stabilize the di

Page 25 - 3.2.3 Measuring Range

<3. Parameter Setting>3-17IM 01C25R02-01E3.3 Detailed Setup3.3.1 Analog Output Signal Adjustable RangeOutput signal adjustable range at n

Page 26 - 3.2.4 Units

<3. Parameter Setting>3-18IM 01C25R02-01EF0304.aiAvailable displays Description and related parameters% of PV range(PV %)Indicates input va

Page 27 - Mode Setup

iiIM 01C25R02-01E3.3.7 Trim Analog Output ...3-223.3.8 External Switch Mode .

Page 28 - 3.3 Detailed Setup

<3. Parameter Setting>3-19IM 01C25R02-01E• Procedure to call up the displayDD (HART 5/7)DTM (HART 7)[Root Menu] → Detailed setup → Display

Page 29

<3. Parameter Setting>3-20IM 01C25R02-01ETypical Unit Conversion FactorUse “kg” in case of mass owUse “m3” in case of volume owUse “Nm3”

Page 30 - 3.3.5 Total Flow Setup

<3. Parameter Setting>3-21IM 01C25R02-01E(1) Auto Sensor TrimApplying reference pressure of 0% and 100% of the measurement range to the tr

Page 31 - 3.3.6 Sensor Trim

<3. Parameter Setting>3-22IM 01C25R02-01E• Procedure to call up the display for differential pressureDD (HART 5/7)DTM (HART 7)[Root Menu]

Page 32

<3. Parameter Setting>3-23IM 01C25R02-01E3.3.9 CPU Failure Burnout Direction and Hardware Write ProtectThere are two slide switches on the

Page 33 - 3.3.8 External Switch Mode

<3. Parameter Setting>3-24IM 01C25R02-01E(1) Alarm SettingSelect the process variable at Process Alert which the alarm is set, then set th

Page 34 - 3.3.11 Alarm

<3. Parameter Setting>3-25IM 01C25R02-01E• Procedure to call up the displayDD (HART 5/7)DTM (HART 7)[Root Menu] → Detailed setup →Output c

Page 35

<3. Parameter Setting>3-26IM 01C25R02-01EExample of Pulse OutputF0307.aiT = 1secThe number of pulses (PPS)(*1) Example of scaled pulse Pul

Page 36

<3. Parameter Setting>3-27IM 01C25R02-01E• Procedure to call up the displayDTM (HART 5) Diag and Service → Service → Loop test →Display It

Page 37 - Pulse Rate =10kg

<3. Parameter Setting>3-28IM 01C25R02-01ENOTEThe output process value while simulation can be monitored by LCD and through communication as

Page 38

iiiIM 01C25R02-01EAppendix 1. Safety Instrumented Systems Installation ...A1-1A1.1 Scope and Purpose ...

Page 39

<3. Parameter Setting>3-29IM 01C25R02-01E(5) Squawk (Effective only when setting to HART 7)This feature can be used to identify the commun

Page 40

<3. Parameter Setting>3-30IM 01C25R02-01ETable 3.2 Basic Flow Calc Set-up StepsStep Parameter Contents Remarks1 Flow Calc Mode Select Basi

Page 41

<3. Parameter Setting>3-31IM 01C25R02-01EFlow unit categoryTable 3.5 Mass Flow UnitUnit LCD Communicationgrams per second g/s ←grams per m

Page 42 - Flow unit category

<3. Parameter Setting>3-32IM 01C25R02-01E3.3.15 Burst Mode3.3.15.1 In the case of using HART 5When the Burst mode is enabled, the transmitt

Page 43 - 3.3.15 Burst Mode

<3. Parameter Setting>3-33IM 01C25R02-01ETable 3.8 Burst parametersCommand parameter Burst CommandBurst Msg Trigger ModeBurst Trigger Sour

Page 44

<3. Parameter Setting>3-34IM 01C25R02-01E(2) Burst mode setting procedure• Procedure to call up the displayDD (HART 7)DTM (HART 7)[Root M

Page 45

<3. Parameter Setting>3-35IM 01C25R02-01Ea. Burst CommandSelect the transmission data at Burst Command parameter.Burst Command Command par

Page 46

<3. Parameter Setting>3-36IM 01C25R02-01Ee. Burst ModeWhen the Burst mode is set to “Wired HART Enabled”, the transmitter starts to send t

Page 47

<3. Parameter Setting>3-37IM 01C25R02-01Eb) Acknowledge Event NoticationExecute Acknowledge Event Notication method.1) Set to Enter Even

Page 48 - 3.3.16 Multidrop Mode

<3. Parameter Setting>3-38IM 01C25R02-01E3.3.16.2 In the case of using HART 7“Multidropping” transmitters refer to the connection of severa

Page 49

<1. Introduction>1-1IM 01C25R02-01E1. IntroductionThank you for purchasing the DPharp EJX multivariable transmitter.EJX multivariable tran

Page 50

<3. Parameter Setting>3-39IM 01C25R02-01E4) Applying the new protocol revisiona. Close the conguration tool After completion of Chg uni

Page 51 - 4. Diagnostics

<4. Diagnostics>4-1IM 01C25R02-01E4. Diagnostics4.1 Self-Diagnostics4.1.1 Identify Problems by Using the HART Conguration ToolThe HART

Page 52

<4. Diagnostics>4-2IM 01C25R02-01E4.1.2 Checking with Integral IndicatorNOTEIf an error is detected by running self-diagnostics, an error

Page 53 - 4.2 Advanced Diagnostics

<4. Diagnostics>4-3IM 01C25R02-01ENOTETime Stamp is reset when powering on.(5) Conguration Change CounterThe Conguration Change Counter

Page 54

<4. Diagnostics>4-4IM 01C25R02-01EIMPORTANT• The pressure uctuation amplitude in uids must be sufciently large for blockages to be dete

Page 55 - EJX multivariable transmitter

<4. Diagnostics>4-5IM 01C25R02-01E Functional block diagramThe gure below shows the functional block diagram of ILBD.EJX multivariable t

Page 56 -  A/B Blocking Detection

<4. Diagnostics>4-6IM 01C25R02-01E4.2.2.1 Blockage Detection Limit parameterWhen the parameter based on pressure uctuation exceeds the

Page 57 -  H Side Blocking Detection

<4. Diagnostics>4-7IM 01C25R02-01E H/L Blocking DetectionEJX differential pressure transmitter enables to detect both-, a high-, or low-p

Page 58 - Diag Applicable

<4. Diagnostics>4-8IM 01C25R02-01ENOTE• Ref fDP must be larger than the specied level shown in Table 4.4 (refer to subsection 4.2.2.6). N

Page 59 -  Diag Supp Count

<4. Diagnostics>4-9IM 01C25R02-01E4.2.2.3 Operation Parameters Diag ModeDiag Mode gives the directive for the ILBD operation. There are

Page 60 - <4. Diagnostics>

<1. Introduction>1-2IM 01C25R02-01E1.1 Safe Use of This Product For the safety of the operator and to protect the instrument and the syste

Page 61 - 4.2.2.4 Operating Procedure

<4. Diagnostics>4-10IM 01C25R02-01EBAAAB BBAF0406.aiDiag Supp Count (Number of times: 3)Diag Supp Count (Number of times: 3)Diag PeriodDiag

Page 62 -  Alarm Status

<4. Diagnostics>4-11IM 01C25R02-01E4.2.2.4 Operating ProcedureThe basic ow of the ILBD operation is as follows.1) Initial setting2) Condi

Page 63

<4. Diagnostics>4-12IM 01C25R02-01EBitDD (HART 5/HART 7) DTM (HART 7)DTM (HART 5) Category0Not used Not used1Not used Not used2A Blocking

Page 64

<4. Diagnostics>4-13IM 01C25R02-01E<Example>When the level range that can be measured by the transmitter with 100 kPa span is –80 to

Page 65

<4. Diagnostics>4-14IM 01C25R02-01E• Procedure to call up the displayDD (HART 5/7)DTM (HART 7)[Root Menu] → Diag/Service → Diag Parameters

Page 66

<4. Diagnostics>4-15IM 01C25R02-01E fDP is not enough.No blockage can be detected if fDP is not larger than the specied value. Only fD

Page 67 -  Threshold Value

<4. Diagnostics>4-16IM 01C25R02-01E Invalid Ref F, SPH, SPL, or DPWhen the enough reference uctuation value is not obtained, an alarm of

Page 68 -  Diag DPComp

<4. Diagnostics>4-17IM 01C25R02-01E4.2.2.10 TuningWhen the pressure uctuation amplitude in uids is not sufciently large or an alarm is o

Page 69 - 4.2.2.12 ILBD Parameter List

<4. Diagnostics>4-18IM 01C25R02-01ECombination of threshold for sensitivity parameterHigh Medium LowLim fDPmax 1.50 3.00 3.00Lim fDPmin 0.4

Page 70

<4. Diagnostics>4-19IM 01C25R02-01E4.2.2.12 ILBD Parameter List# Parameter name Default value Explanation1 Diag Error 0x0000 The results de

Page 71 - 4.2.3 Heat Trace Monitoring

<1. Introduction>1-3IM 01C25R02-01E1.3 ATEX DocumentationThis is only applicable to the countries in European Union.GBDKIENLSFPFDSLTLVPLES

Page 72

<4. Diagnostics>4-20IM 01C25R02-01E# Parameter name Default value Explanation28 Ref fSPl Value of fSPl obtained under normal condition.29 R

Page 73

<4. Diagnostics>4-21IM 01C25R02-01E4.2.3 Heat Trace MonitoringThe EJX multivariable transmitter with Heat trace monitoring function calcul

Page 74

<4. Diagnostics>4-22IM 01C25R02-01EAmplifier TemperatureCapsule TemperatureAmp tempCap tempFlg temp CoefFlg tempExecution HTMCT+(CT-AT) x C

Page 75

<4. Diagnostics>4-23IM 01C25R02-01E4.3 Alarms and CountermeasuresTable 4.5 Alarm Message SummaryIntegral indicator HART conguration too

Page 76

<4. Diagnostics>4-24IM 01C25R02-01EIntegral indicator HART conguration tool display Cause4-20mA Output operation during error Counterm

Page 77 - Conguration Changed Flag

<4. Diagnostics>4-25IM 01C25R02-01EIntegral indicator HART conguration tool display Cause4-20mA Output operation during error Counterm

Page 78

<4. Diagnostics>4-26IM 01C25R02-01EIntegral indicator HART conguration tool display Cause4-20mA Output operation during error Counterm

Page 79

<4. Diagnostics>4-27IM 01C25R02-01ETable 4.6 HART Conguration Tool Error MessagesError message Probable cause CountermeasureInvalid selec

Page 80

<4. Diagnostics>4-28IM 01C25R02-01ETable 4.10 Relationship between Alarm and Status available for HART 7Integral IndicatorHART conguratio

Page 81

<4. Diagnostics>4-29IM 01C25R02-01EIntegral IndicatorHART conguration tool displayDevice StatusExtended Device StatusValue and Status (Da

Page 82

<2. Connection>2-1IM 01C25R02-01E2. Connection2.1 Integral Indicator Display When Powering OnFor models with the integral indicator code

Page 83 - 5. Parameter Summary

<4. Diagnostics>4-30IM 01C25R02-01EIntegral IndicatorHART conguration tool displayDevice StatusExtended Device StatusValue and Status (Da

Page 84

<4. Diagnostics>4-31IM 01C25R02-01EIntegral IndicatorHART conguration tool displayDevice StatusExtended Device StatusValue and Status (Da

Page 85

<4. Diagnostics>4-32IM 01C25R02-01EIntegral IndicatorHART conguration tool displayDevice StatusExtended Device StatusValue and Status (Da

Page 86

<5. Parameter Summary>5-1IM 01C25R02-01E5. Parameter SummaryFunction Label Item Contents Default value Handling *1Analog output AO alm typ

Page 87

<5. Parameter Summary>5-2IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Device informationCountryHART 7Country code U

Page 88

<5. Parameter Summary>5-3IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Diag Lim Lim fDPmax Upper limit for Ratio fD

Page 89

<5. Parameter Summary>5-4IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Display setup Bar Indicator Bar indicator Off

Page 90

<5. Parameter Summary>5-5IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Event NoticationEvent Notication ControlHAR

Page 91

<5. Parameter Summary>5-6IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Event NoticationLatched Device StatusHART 7D

Page 92

<5. Parameter Summary>5-7IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Master test Master test Software reset and se

Page 93

<2. Connection>2-2IM 01C25R02-01E(1) Conrmation by the name plate The HART protocol revision is shown by the last number of the serial n

Page 94 - Installation

<5. Parameter Summary>5-8IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Process variablesEngr exp Exponents x1, x10,

Page 95

<5. Parameter Summary>5-9IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Process variablesQV isHART 7Current QV RQV Da

Page 96 - Abbreviations

<5. Parameter Summary>5-10IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Self test Self test Self-diagnostics MSensor

Page 97 - Appendix 2. ILBD Check List

<5. Parameter Summary>5-11IM 01C25R02-01EFunction Label Item Contents Default value Handling *1Status Status group 1 Device status informat

Page 98 - Checklist (2/5)

<Appendix1. Safety Instrumented Systems Installation>A1-1IM 01C25R02-01EAppendix 1. Safety Instrumented Systems InstallationWARNING The co

Page 99 - Checklist (3/5)

<Appendix1. Safety Instrumented Systems Installation>A1-2IM 01C25R02-01ETable A1.2 Proof TestingTesting method Tools required Expected out

Page 100 - Checklist (4/5)

<Appendix1. Safety Instrumented Systems Installation>A1-3IM 01C25R02-01EA1.3 Denitions and AbbreviationsA1.3.1 DenitionsSafety Freedom

Page 101 - Checklist (5/5)

<Appendix2. ILBD Check List>A2-1IM 01C25R02-01EAppendix 2. ILBD Check ListFill out the below checklist according to the operation ow of t

Page 102 - Revision Information

<Appendix2. ILBD Check List>A2-2IM 01C25R02-01EChecklist (2/5)No. Items Parameters Result Example8 Alarm status• Check the alarm status sh

Page 103 - Blank Page

<Appendix2. ILBD Check List>A2-3IM 01C25R02-01EChecklist (3/5)Go to the following step according to the result of "Invalid Ref xx"