IM 01E20D01-01E
5-3
5. MAINTENANCE
(6) Move aside the electrode lead wire and avoiding
the screw, insert end A of the tool to the electrode
and loosen the electrode holder using the torque
driver.
F0506.EPS
Torque driver
End A
(7) Screw end B of the tool into the screw block of the
electrode, and pull the tool straight out to remove
the electrode holder.
F0507.EPS
End B
5.2.2 Cleaning Replaceable Elec-
trodes
(1) Clean the electrode surface (wetted part) with
alcohol or other cleaning agents.
O-ring
Electrode surface
Electrode holde
F0508.EPS
NOTE
The screw threads are coated with an anti-
seizing compound. Do not let the compound
come in contact with the electrode.
(2) Clean the surface of the lining seal with an alcohol
moistened swab.